I.N. Posokhov1, O.S. Orlova1,2, E.А. Sukhodolova2, A.V. Lystseva2

Validation of an Automatic Sphygmomanometer with Cuff Range of “17-22”
for Measurements in Adults with Upper Arm Circumference of 22 Centimeters or Less

1International Office, State Research Center – Burnasyan Federal Medical Biophysical Center of Federal Medical Biological Agency, Moscow, Russiа

2Pirogov Russian National Research Medical University, Moscow, Russia

Contact person: Lystseva Aleksandra Vasilevna: aleksandra.lyst@gmail.com

 

Abstract
To validate a blood pressure (BP) monitor according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) in a special population of adults with an upper arm circumference of 22 cm or less 37 subjects were recruited. The previously validated in general population OMRON M2 Basic (HEM-7121) was used with a HEM CS24 cuff designed for an upper arm circumference of 17–22 cm.  Sequential measurements of BP were obtained using the test device and the standard mercury sphygmomanometer. A total of 108 sets of comparison data were obtained and analyzed. According to the validation criterion 1, the mean ± SD of the differences was -7.81 ± 5.20 / -10.66 ± 5.48 mmHg. In this special group HEM-7121 and HEM CS24 do not meet the requirements of ISO 81060-2:2018.

Keywords: AAMI/ESH/ISO 81060-2:2018, blood pressure monitor, blood pressure measurement, hypertension, validation

For citation: Posokhov IN, Orlova OS, Sukhodolova EА, Lystseva AV. Validation of an Automatic Sphygmomanometer with Cuff Range of “17-22” for Measurements in Adults with Upper Arm Circumference of 22 Centimeters or Less A.I. Burnasyan Federal Medical Biophysical Center Clinical Bulletin. 2024.2:53-56. (In Russian) DOI: 10.33266/2782-6430-2024-2-53-56

 

REFERENCES

  1. Посохов И.Н., Праскурничий Е.А., Орлова О.С., Суходолова Е.А., Лысцева А.В., Лобова М.Д. Стандарты проверки точности автоматических манжеточных сфигмоманометров: современное состояние дел // Клинический вестник ФМБЦ им. А.И. Бурназяна. 2024. № 1. С. 5–13. [Posokhov I.N., Praskurnichiy E.A., Orlova O.S., Sukhodolova E.А., Lystseva A.V., Lobova M.D. Validation Standards for Automatic Cuff-Based Sphygmomanometers: State of the Art. Klinicheskiy vestnik FMBTS im. A.I. Burnazyana  = A.I. Burnasyan Federal Medical Biophysical Center Clinical Bulletin. 2024;1:5–13 (In Russ.)]. DOI: 10.33266/2782-6430-2024-1-05-13.
  2. Stergiou G.S., Alpert B.S., Mieke S., Wang J., O’Brien E. Validation Protocols for Blood Pressure Measuring Devices in the 21st Century. J. Clin. Hypertens (Greenwich). 2018;20;7:1096-1099. doi: 10.1111/jch.13294.
  3. International Organization for Standardization. ISO 81060-2:2018. Noninvasive Sphygmomanometers: Part 2: Clinical Investigation of Intermittent Automated Measurement Type. 2018. URL: https://www.iso.org/standard/73339.html. (Accessed 14.03.2024).
  4. Stergiou G.S., Menti A., Asayama K., et al. Accuracy of Automated Cuff Blood Pressure Monitors in Special Populations: International Organization for Standardization (ISO) Task Group Report and call for research. J. Hypertens. 2023;41;5:811-818. doi: 10.1097/HJH.0000000000003403.
  5. Takahashi H., Yoshika M., Yokoi T. Validation of Three Automatic Devices for the Self-Measurement of Blood Pressure According to the European Society of Hypertension International Protocol Revision 2010: the Omron HEM-7130, HEM-7320F, and HEM-7500F. Blood Press Monit. 2015;20;2:92-97. doi: 10.1097/MBP.0000000000000096.
  6. O’Sullivan J.J., Derrick G., Griggs P.E., Wren C. Validation of the Takeda 2421 Ambulatory Blood Pressure Monitor in Children. J. Med. Eng. Technol. 1998;22;3:101-105. doi: 10.3109/03091909809062474.
  7. Belsha C.W., Wells T.G., Bowe Rice H., Neaville W.A., Berry P.L. Accuracy of the SpaceLabs 90207 Ambulatory Blood Pressure Monitor in Children and Adolescents. Blood Press Monit. 1996;1;2:127-133.
  8. Ostchega Y., Nwankwo T., Sorlie P.D., Wolz M., Zipf G. Assessing the Validity of the Omron HEM-907XL Oscillometric Blood Pressure Measurement Device in a National Survey Environment. J. Clin. Hypertens (Greenwich). 2010;12;1:22-28. doi: 10.1111/j.1751-7176.2009.00199.x.

Conflict of interest. The authors declare no conflict of interest.
Financing. The study had no sponsorship. 
Contribution. Article was prepared with equal participation of the authors.
Article received: 20.04.2024. Accepted for publication: 13.05.2024