A.I. Burnasyan FMBC clinical bulletin. 2024 № 2
I.N. Posokhov1, O.S. Orlova1,2, A.V. Lystseva2, E.А. Sukhodolova2
Validation of an Automatic Sphygmomanometer with “360°” Cuff for Measurements in
Adults with Upper Arm Circumference of 22 Centimeters or Less
1International Office, State Research Center – Burnasyan Federal Medical Biophysical Center of Federal Medical Biological Agency, Moscow, Russiа
2Pirogov Russian National Research Medical University, Moscow, Russia
Contact person: Sukhodolova Ekaterina Arsenyevna: prosto0600788@gmail.com
Abstract
To validate a blood pressure (BP) monitor according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) in a special population of adults with an upper arm circumference of 22 cm or less 37 subjects were recruited. Sphygmomanometer “Armed” identical to previously validated in the general population group YuWell YE660E was used with a “360°” cuff designed for an upper arm circumference of 18 – 36 cm. Sequential measurements of BP were obtained using the test device and the standard mercury sphygmomanometer. A total of 108 sets of comparison data were obtained and analyzed. According to the validation criterion 1, the mean ± SD of the differences was 8.00 ± 6.30 / -16.11 ± 5.15 mmHg. In this special group “Armed” YE660E with “360°” cuff do not meet the requirements of ISO 81060-2:2018.
Keywords: AAMI/ESH/ISO 81060-2:2018, blood pressure monitor, blood pressure measurement, hypertension, validation
For citation: Posokhov I.N., Orlova O.S., Lystseva A.V., Sukhodolova E.А. Validation of an Automatic Sphygmomanometer with “360°” Cuff for Measurements in Adults with Upper Arm Circumference of 22 Centimeters or Less. A.I. Burnasyan Federal Medical Biophysical Center Clinical Bulletin. 2024.2:57-60. (In Russian) DOI: 10.33266/2782-6430-2024-2-57-60
REFERENCES
- Посохов И.Н., Праскурничий Е.А., Орлова О.С., Суходолова Е.А., Лысцева А.В., Лобова М.Д. Стандарты проверки точности автоматических манжеточных сфигмоманометров: современное состояние дел // Клинический вестник ФМБЦ им. А.И. Бурназяна. 2024. № 1. С. 5–13. [Posokhov I.N., Praskurnichiy E.A., Orlova O.S., Sukhodolova E.А., Lystseva A.V., Lobova M.D. Validation Standards for Automatic Cuff-Based Sphygmomanometers: State of the Art. Klinicheskiy vestnik FMBTS im. A.I. Burnazyana = A.I. Burnasyan Federal Medical Biophysical Center Clinical Bulletin. 2024;1:5–13 (In Russ.)]. DOI: 10.33266/2782-6430-2024-1-05-13.
- Stergiou G.S., Alpert B.S., Mieke S., Wang J., O’Brien E. Validation Protocols for Blood Pressure Measuring Devices in the 21st Century. J. Clin. Hypertens (Greenwich). 2018;20;7:1096-1099. doi: 10.1111/jch.13294.
- International Organization for Standardization. ISO 81060-2:2018. Noninvasive Sphygmomanometers: Part 2: Clinical Investigation of Intermittent Automated Measurement Type. 2018. URL: https://www.iso.org/standard/73339.html. (Accessed 14.03.2024).
- Stergiou G.S., Menti A., Asayama K., et al. Accuracy of Automated Cuff Blood Pressure Monitors in Special Populations: International Organization for Standardization (ISO) Task Group Report and call for research. J. Hypertens. 2023;41;5:811-818. doi: 10.1097/HJH.0000000000003403.
- Chen Q, Lei L, Li Y, Wang JG. Validation of the YuWell YE690A upper-arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit. 2017 Oct;22(5):295-297. doi: 10.1097/MBP.0000000000000278.
- O’Sullivan J.J., Derrick G., Griggs P.E., Wren C. Validation of the Takeda 2421 Ambulatory Blood Pressure Monitor in Children. J. Med. Eng. Technol. 1998;22;3:101-105. doi: 10.3109/03091909809062474.
- Belsha C.W., Wells T.G., Bowe Rice H., Neaville W.A., Berry P.L. Accuracy of the SpaceLabs 90207 Ambulatory Blood Pressure Monitor in Children and Adolescents. Blood Press Monit. 1996;1;2:127-133.
- Ostchega Y., Nwankwo T., Sorlie P.D., Wolz M., Zipf G. Assessing the Validity of the Omron HEM-907XL Oscillometric Blood Pressure Measurement Device in a National Survey Environment. J. Clin. Hypertens (Greenwich). 2010;12;1:22-28. doi: 10.1111/j.1751-7176.2009.00199.x.
Conflict of interest. The authors declare no conflict of interest.
Financing. The study had no sponsorship.
Contribution. Article was prepared with equal participation of the authors.
Article received: 20.04.2024. Accepted for publication: 13.05.2024